The NMP-1000 is an analog, wide range, linear current mode module with range switching. Ranges may be selected either manually or automatically, locally or remotely. Full scale sensitivities from 1 x 10-11 to 1 x 10-3 A are provided with indication of percent power on a front panel meter, as well as a front panel LED indication of each decade range. For remote operation, a separate panel assembly can be provided with identical range indicating LEDs, display of current and voltage, as well as channel self test functions. Compensation voltage as well as adjustable high voltage power supplies are provided in the module to allow operation with most existing current mode detectors.



10-11 to 10-4 A or 10-10 A to 10-3 A in seven one decade ranges

Self-bailing illuminated pushbutton or automatic range switching (manual or auto operator selected). Provision for remote range switching

LINEARITY (LOG CONFORMITY) + 1% OF THE LINEAR ± 1.0% of full scale upper 5 ranges; ±2.5% of full scale lower 2 ranges
TEMPERATURE COEFFICIENT ± 0.15% per °C maximum over 10° to 55°C
CALIBRATION/TEST 2 fixed currents for calibration, 1 adjustable current for multi-range function test and trip testing. HV trip test, test modes selected sequentially by front panel control
RESPONSE TIME CONSTANTS 1 msec 10-8 to 10-3 A, 10 msec 10-9 to 10-8 A; 100 msec to 10-11 to 10-9
Remote Meter: 0–10 V full scale Recorder: 0–1.0 V full scale
Remote Meter: 0–1 mA full scale Optional: 4.0–20 mA full scale
Recorder: 0–0.1 V full scale High Voltage: +300 to +800 VDC @2.6W
  Compensation: 0 to–150VDC
High voltage Adjustable with
High power level +5 V logic level
Low power level and two form C
Period contacts per trip
POWER REQUIRED 117 VAC ± 10% 50/60 Hz @ 1.0 A